ZENO
®
-3200 USER MANUAL
Coastal Environmental Systems (206) 682-6048
Page B-6
12: COM1, COM2, COM3
13,17,18: COM2, COM3
14: CH10 through CH16 (uses two inputs)
15,16: COM1, COM2 or COM3
4 Analog Channel Gain 1, 10 or 100
5 Analog Channel Attenuation 1, 2, 4 or 10
Section 10.3.1
6 Switched Power Code
0 - no power
1 - 5 V switched “A”
2 - 5 V switched “B”
3 - 12 V switched “A”
4 - 12 V switched “B”
5 - 12 V switched “C”
Section
10.3.10
7 Sensor Excitation Voltage Code
0 - no excitation voltage
1 - 1.25 V DC
2 - 2.50 V DC
3 - 5.00 V DC
8 Switched Excitation Return Codes
0 - no excitation voltage
A, B, C, D, E
Section
10.3.12
9 Switched Power Warm-Up Time
0 to 255 seconds Section
10.3.11
10 Sensor Sample Count 1 to 255 Section 10.3.9
11 Maximum Sensor Readings
0 (measurements throughout sample duration),
1 to 32767
Section 10.3.8
12 Sensor Timing Loop 1 to 4 Section 10.3.7
13-
15
Conversion Coefficients A, B, C
-3.410
38
to 3.410
38
Section
10.3.13
16 Retry Count 0 to 7
17 (Depends upon sensor type) (Range depends upon sensor type)
18 Sensor Baud Rate 300 to 19200 baud
19 Sensor Port Type (range depends upon sensor type)
20 Sensor Bits Per Character 5 to 8
21 Sensor Parity N (none), E (even) or O (odd)
22 Sensor Start Bits 1 or 2
23 Sensor Stop Bits 1 or 2
24 Sensor GSI Script Number 1 to 6
Section 10.3.5
25 Sensor Data Collection Control S0.1 or reference to process output Section 12.2.2
P PROCESS MENU
1 Process Category Code
1 - General
2 - Environmental
3 - Alarm
4 - Special Sensors
5 - Arithmetic
6 - Output and Control
Section
10.4.1
Appendix D
2 Process Type Code
(Values dependent on process category
code)
1 - General Processes Category
1 - Immediate Data
2 - Average (over Sample Time)
3 - Running Average
4 - Block Average
5 - Accumulate Data (over Sample
Time)
6 - Rate of Change
Section
10.4.1
Appendix D
Comentarios a estos manuales