Campbell SAT ARGOS Manual de usuario Pagina 226

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 394
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 225
ZENO
®
-3200 USER MANUAL
Coastal Environmental Systems (206) 682-6048
Page 218
There are four Sensor Timing Loops. Each Sensor Timing Loop defines
how often the ZENO
®
-3200 takes a reading from the sensor (or sensors)
utilizing the Sensor Timing Loop during the Sample Duration.
The time defined for each of the four Sensor Timing Loops is in the Sensor Timing Loop Menu, which is
accessible from the Zeno Program Menu and will look something like this:
SENSOR TIMING LOOP MENU
(Cn/m) Change Item n To Value m (H) Help
(Z) Zeno Program Menu
Item 1: 0.5 (Timing Loop #1 Period)
Item 2: 1.0 (Timing Loop #2 Period)
Item 3: 1.5 (Timing Loop #3 Period)
Item 4: 2.0 (Timing Loop #4 Period)
>
As an example, configure the Sample Interval to be 20 seconds, the Sample Duration to be 19 seconds
and the Sensor Timing Loop period to be 2.5 seconds. The ZENO
®
-3200 would operate according to the
following timing diagram:
Sample Interval |---------------------------------------| 20s
Sample Duration |-------------------------------------| | 19s
Sensor Timing Loop ^ ^ ^ ^ ^ ^ ^ ^ 2.5s
Sensor Reading at each ^
Each sensor type requires a certain amount of time to collect data. From the above example, let's say that
the sensor requires up to 1 second to take a measurement. The addition of the time required by the sensor,
to the timing diagram, results in the following:
Sample Interval |---------------------------------------| 20s
Sample Duration |-------------------------------------| | 19s
Sensor Timing Loop ^- ^- ^- ^- ^- ^- ^- ^- 2.5s
Sensor Reading at each ^
Sensor reading complete at each
Vista de pagina 225
1 2 ... 221 222 223 224 225 226 227 228 229 230 231 ... 393 394

Comentarios a estos manuales

Sin comentarios