ZENO
®
-3200 USER MANUAL
Coastal Environmental Systems (206) 682-6048
Page 19
> c4/2
SYSTEM FUNCTIONS MENU
(Cn/m) Change Item n To Value m (I) Contact Information
(S) System Date and Time (E) Save Parameters To EEPROM
(T) Calibrate Internal Temperature (U) User Menu
(V) Program Version (Q) Quit
(K) Constants Menu (H) Help
(B) BIT Names Menu
Item 1: 0 (Primary Unit/Experiment ID)
Item 2: 0 (Secondary Unit/Experiment ID)
Item 3: 1 (Data Dump Format)
Item 4: 2 (Real Time Output Format)
Item 5: 0 (Add Compass To Vane)
Item 6: 0 (Compass Offset)
Item 7: 0 (Barometer Elevation)
Item 8: (Bad Sensor Value Replace)
>
You can now exit the User Interface by typing Q{ENTER}. Once you see the next real-time data
messages, they will now be space delimited. Your screen should look something like this:
> q
Exiting user interface.
01/07/12 10:01:55 22.8 13.8
01/07/12 10:02:55 22.8 13.8
01/07/12 10:03:55 22.8 13.8
01/07/12 10:04:55 22.8 13.8
01/07/12 10:05:55 22.8 13.8
An important note about space delimiting: When you want to use space delimiting, it is important that
the field widths for each data value are set to a sufficiently large number. This number will allow the full
range of the data value to be contained and provide at least one space between it and the preceding data
value in the message. Otherwise, the data values will run together with no space in between.
19
The change you made is currently residing in the ZENO
®
-3200’s RAM only. It will be effective until the
ZENO
®
-3200 powers down. When the ZENO
®
-3200 powers up again, the settings in EEPROM will be
read into RAM again and your change will be lost. If you want configuration changes to be maintained
after the ZENO
®
-3200 powers down, you must save the modified configuration into EEPROM so that the
changes will be restored the next time the ZENO
®
-3200 is turned on.
When you want to save a modified configuration in case you power down the
ZENO
®
-3200, type E{ENTER} from any menu that offers the "Save
Parameters to EEPROM" option.
20
19
Setting the field widths for each data value is done in the Data Output Menu, which is part of the Zeno Program
Menu. This is discussed in Section 10.6.6.
20
Refer to Section 2.3.2.