Example 2: Temperature and Water Pressure Measurements
Writing a data logger program
1. Proceed as in the previous example to set up the program using SCWin following steps
one and two above. Choose an appropriate name for the new program.
3. From the home page press “Sensors”. Under “Available Programs” open the “Generic
Measurements” folder. Select “Full Bridge”, the transducer is of this type of circuit,
and press the red arrow, or double click it (See Fig. 10). Set the “Range of Sensor
Voltage” to +/- 250 mv (See Fig. 11). Then click OK. The sensor should now be
listed under “Selected Sensors” on the left side of the screen. Again under available
sensors, go to temperature and select the Type T thermocouple as described previously.
After the thermocouple has been selected both sensors should be listed under “Selected
Sensors”.
Fig. 10 Defining Transducer
4. Press the “Wiring Diagram” button at the bottom of the window as illustrated in Fig.10
and a window depicting the wiring diagram of the data logger for a pressure transducer
and Type T thermocouple appears, see Fig. 12. Connect the blue wire of the
thermocouple with the port 2H and the red wire of the thermocouple with the port 2L of
the CR10X data logger (you will need a small flathead screw driver to tighten the
thermocouple wire down).