
Section 6. Calibration
9. Weigh 5 to l0 equal increments of the sediment so that the total dry weight
will produce the maximum concentration expected at the monitoring site.
10. For each sediment standard, repeat Steps 2 through 4 of Section 6.1.3.
11. After all sediment values have been logged, follow Steps 6 through 9 of
Section 6.1.3 to complete the calibration.
6.3 Salinity, Pressure and Temperature
Calibrations
Due to the specialized equipment involved for salinity, pressure and
temperature calibration, it is recommended that the instrument be returned to
Campbell Scientific, Inc. if any of these sensors are not operating with
specified accuracy.
6-7
Comentarios a estos manuales