Section 6. Calibration
There is no way to cancel a calibration after the
Compute button is clicked.
WARNING
Review the quality assurance checks discussed later in this section. Look at the
plot of residuals, which show the differences between the standard and
computed NTU values. The average residual magnitude should be less than
1% of the calibration range. For example, a calibration range of 2000 NTU
(maximum minus minimum NTU values) should produce an average residual
less than 20 NTU. Residuals higher than 1% of the calibration range indicate
that errors were made in the procedure. By inspecting the plot of residuals, you
can decide which values to repeat, edit, or delete. Repeat values using the
procedures described above. To edit a value,
6. Click the Edit button; select the number of the calibration value you want
to change from the data table; enter the revised calibration value; and click
OK.
7. The data table and plot will be updated. Use the Delete button to remove a
calibration value by number from the data table.
8. Once satisfied with the calibration, click the Compute button and make a
final quality check.
9. Use the Done / Exit button to return to the OBS-3A Utility.
6.1.4 Making Turbidity Standards
1. For the zero NTU calibration point you will need a black 20”X14”X16”
container filled with clean tap water. A Rubbermaid plastic storage box
makes a suitable container.
2. For calibrations from 50-250 NTU use a 6” diameter container with 2
liters of filtered water or a premixed standard. Hold the sensor in the
container so that the beam looks down and across the diameter to prevent
the beam from reflecting off the wall.
3. A 4” diameter container can be used for solutions of 250 NTUs. Add one
liter of filtered water and the necessary amount of formazin. See TABLE
6.1-1 or the formula below for the preparation of standards.
4. Stir the formazin solution with the spoon and position the OBS-3A in the
standard.
5. Go to Step 2 of OBS-3A Utility Steps (Section 6.1.3).
The formula for preparing turbidity standards other than shown in TABLE
6.1-1 is:
»
¼
º
«
¬
ª
stkdw
stk
stkstd
VV
V
TT
or
»
¼
º
«
¬
ª
u
stdstk
dwstd
stk
TT
VT
V
6-3