Section 5. Maintenance
carefully inspected every six months and serviced before all deployments
longer than one month.
1. Disassemble O-ring seals and inspect mating surfaces for pits and
scratches.
2. Inspect O-rings for cuts and nicks; replace if necessary using spares
provided.
3. Clean O-rings and mating surfaces with a cotton swab and alcohol.
Remove fibers from groove and mating surfaces then grease O-rings with
DOW Compound 55 and reassemble.
5.6 Antifoulant Coatings
Clear TBTA antifoulant coating or toxin-impregnated collars can be used for
monitoring in biologically active waters. TBTA prevents most marine algae
and encrusting animals from growing on optical surfaces for up to two months
with minimal loss of IR transmission. It is illegal to use TBTA in many places
so check applicable water quality regulations in your area before using TBTA
coated OBS sensors. Use of TBTA is the sole responsibility of the user
.
5.7 User-serviceable Parts
See Section 10—Specifications for a list of user-serviceable parts.