Section 3. Operations
box to select ONLY change host computer port. Then click Apply and the
button.
If you get the OBS-3A information box, the baud rate of the unit is
synchronized with your PC. If you don’t get an information box, repeat the
above procedure.
FIGURE 3.4-1. Dialog Box for Changing Baud Rate
3.5 Testing Sensors
1. Before daily operations and deployments, verify the instrument works by
clicking
Open Plot, and then clicking Survey. Select all
installed sensors and click Start Survey.
2. Wave your hand in front of the OBS sensor; the turbidity signal on the top
plot of FIGURE 3.5-1 will fluctuate and data will scroll.
3. Blow on the temperature sensor to observe an increase in temperature (red
trace on the middle plot of FIGURE 3.5-1).
4. Blow into the pressure sensor and a small elevation in the pressure signal
will occur (bottom plot of FIGURE 3.5-1).
5. Dip the sensor in salty water and conductivity will increase (blue trace on
middle plot).
6. Click
OBS-3A Settings to view time, serial numbers, depth
corrections, and software versions (FIGURE 3.5-2).
3-4