
Section 6. Calibration
6.1.2 Preparation
1. Experience has shown that only three calibration points are needed to get
sub-1% accuracy. It is recommended to do one for clean filtered water, a
midrange value (e.g. 125, 250, 500 or 1000 NTU) and one at the high end
of the desired measurement range (e.g. 250, 500, 1000, or 2000 NTU).
2. Scrub the sensor, container, spoon, and glassware with detergent and water
and rinse everything twice with filtered water.
3. To avoid interference from incandescent and solar IR, perform calibrations
under fluorescent lighting and make field checks in the shade.
4. Start the OBS-3A Utility Software and wake the OBS-3A.
6.1.3 OBS-3A Utility Software Steps
1. Open the calibration dialog box with and click the NTU button.
Follow the instructions to place the OBS-3A in a turbidity standard.
You cannot change the OBS-3A calibration after
exiting the calibration dialog box.
WARNING
2. Enter the standard value in the Value box of the calibration dialog and
click the Record button. If mg/l or ppm are selected, choose a sample
duration from TABLE 6.2-1.
3. After the sample has been logged, the table and plot of calibration data
will be updated. Verify that the calibration value is what you intended it to
be.
4. Prepare the next NTU standard and put the OBS-3A in it. Repeat steps 1
and 2 for all calibration values.
5. After the calibration is complete, click the Compute button and inspect
the plot to verify that it resembles the example shown above.
6-2
Comentarios a estos manuales