Campbell 21X Especificaciones Pagina 22

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 21
Number 4250-2010
Revision 0.1
Date MAY 1996
Page 3 of 37
Electronic contacts cleaning fluid
Medium screwdrivers (flat-blade and Phillips-head)
4.0 METHODS
Handar 540A/570A DCPs should be serviced according to the following schedule:
Prior to installation at a field monitoring site
On an annual schedule
Any time the operation or accuracy of the datalogger appears to be suspect
Factory servicing and calibration of the Handar 540A/570A DCP is required when timing
and performance checks indicate that the DCP is not operating within factory specifications.
This section includes five (5) major subsections:
4.1 Post-Field Inspection and Performance Checks
4.2 Routine Laboratory Servicing
4.3 DCP Programming
4.4 Pre-Field Performance Testing
4.5 Archiving Handar 540A/570A DCP Service Records
4.1 POST-FIELD INSPECTION AND PERFORMANCE CHECKS
When a DCP is returned from a field site, the external and internal physical condition is
visually inspected prior to performing any performance tests or laboratory servicing. If the DCP
is received with the power switch in the "ON" position and there are no loose circuit boards,
disconnected or damaged connectors, or other apparent problems that might affect the operation
of the DCP, performance tests that evaluate DCP timing, A/D converter operation, transmission
power, and the DCP program are performed. Results and comments related to inspection and
performance testing are fully documented on the Post-Field Inspection Checklist - Handar
540A/570A DCP (Figure 4-1).
4.1.1 General Information
RECORD Record the DCP serial number, the site it was received from, and
GENERAL the date it was received. The initials of the technician performing
INFORMATION the inspection should also be recorded.
IDENTIFY Identify the DCP model (Figure 4-2 shows the front panel layout
DCP MODEL of each of the three DCP models used by ARS).
NOTE REASON Note whether the DCP was returned for annual servicing (no
FOR RETURN observed operational problems in the field) or for unscheduled
maintenance (unit malfunctioning). If returned for unscheduled
maintenance, describe the observed field symptoms.