Number 4250-2000
Revision 1.0
Date MAY 1996
Page 6 of 7
Documentation Form. The correct reading for each port is shown,
along with the manufacturers' specified accuracy, on the Campbell
21X Datalogger Servicing Documentation Form.
If the datalogger readings for either CAO port differ from the
specified values by more than ± 5.0 millivolts, return the datalogger
to Campbell Scientific for recalibration.
4.4 PULSE COUNTER CHECK
CONNECT Connect the waveform generator to pulse input channel #1.
GENERATOR TO
FREQUENCY
COUNTER
SETUP Setup the waveform generator for a square wave output with a
WAVEFORM frequency of 1000 Hz and an amplitude of 1 volt(rms).
GENERATOR
RECORD The test program will count pulses from the waveform generator
COUNTS for a period of 10 seconds. Record the number of counts in the
pulse counter channel at storage location 09 ( press
*6 9
on the
datalogger). Based on an input frequency of 1000 Hz, a datalogger
count of 10,000 should be displayed.
If the datalogger reading for the pulse counter channel differs from
the specified value by more than ± 5 counts, return the datalogger to
Campbell Scientific for recalibration.
4.5 PANEL TEMPERATURE CHECK
RECORD Read the ambient temperature in the laboratory with the laboratory
AMBIENT reference thermometer. Record this temperature (
°
C) on the
TEMPERATURE Campbell 21X Datalogger Servicing Documentation Form (Figure
4-1).
RECORD Read the datalogger panel temperature at storage location 10 (press
PANEL
*6 10
on the datalogger) and record the reading on the Campbell
TEMPERATURE 21X Datalogger Servicing Documentation Form.
If the datalogger panel temperature measurement differs from the
laboratory reference thermometer reading by more than ± 1.7
°
C,
return the datalogger to Campbell Scientific for recalibration.
4.6 INTERNAL BATTERY SERVICING
RECORD Read the internal battery voltage at storage location 11 (press
*6 11
BATTERY on the datalogger). Record this reading on the Campbell 21X
VOLTAGE Datalogger Servicing Documentation Form.