RECHARGE If the battery voltage is less than 11.76 volts, connect the
BATTERY datalogger to the battery charger. Recharge the battery for eight (8)
hours.
REPLACE Disconnect the datalogger from the battery charger and recheck the
BATTERY battery voltage (press
*6 11
on the datalogger). If the battery
voltage is still less than 11.76 volts, replace the battery as described
below:
• Turn the power switch
OFF
.
• Remove the two front panel screws and carefully raise the front
panel away from the datalogger case.
• Disconnect the used battery from the charging circuit and remove
from the datalogger case.
• Install a fresh battery. Mark the installation date on the battery.
• Remove the datalogger desiccant packets and replace with two
(2) dry half unit DESI PAK desiccant packets.
• Replace the front panel.
• Turn the power switch
ON
and recheck the battery voltage.
4.7 ARCHIVING DATALOGGER SERVICE RECORDS
All service records for Campbell 21X dataloggers are maintained by the instrument
technician. The records are archived by datalogger serial number in three-ring notebooks located
in the ARS instrumentation laboratory.