Verifying Sensor Data with PC208W
Campbell Scientific’s Windows based Data Logger maintenance program, PC208W can
be used to verify that the sensor data collection is proceeding correctly at your new or
existing MicroMet data site. Here are the steps to look at your sensor data:
! Start PC208W
! From the PC208 floating tool bar select the CONNECT button.
! The CR10X data logger connection window will appear. This window has three
tabs along its bottom edge. Select the TOOLS tab.
! From the Station list on the TOOLS tab, select the data station your are visiting. If
the station you are visiting is not listed, See Setting up a New Station in PC208W,
pg 16.
! Connect the SC929 Campbell Cable between the CR10X CS I/O connector and you
PC’s serial data (COM) port.
! Press the CONNECT button on the TOOLS tab. This will open a connection
between your PC and the CR10X.
! Select the NUMERIC DISPLAY tab.
! Numeric values will be displayed on this tab for all the physical sensors installed at
the site. These values are displayed in real numbers that should make sense
compared to actual observed values at the site. If they don’t you should tale a close
look at your sensors, sensor wiring, etc. If all sensors are not displayed on the
numeric display, you can ADD them by pressing the ADD button and dragging them
from the Inloc list which will pop up. We try to place all sensors at the beginning of
the inloc list.