Campbell AVW1 Especificaciones Pagina 37

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 50
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 36
AM16/32B Relay Analog Multiplexer
"4 X 16" Mode
COM H (ODD)
COM L
COM H (EVEN)
COM L
ODD H
ODD L
EVEN H
EVEN L
SENSOR SHIELDS
COM
"4 X 16" Mode
COM H (ODD)
COM L
ODD H
ODD L
/
0
CR23X/
CR800/
CR850/
CR1000
21X/
CR7
CR10(X)
CR3000/
CR5000
AG
EX E E VX
H H H H
L L L L
H H H H
L L L L
G
FIGURE 17. Full Bridge Measurement with Excitation Compensation
6.6 Thermocouple Measurement
The datalogger manuals contain thorough discussions of
thermocouple measurement and error analysis. These topics will not
be covered here.
6.6.1 Measurement Considerations
Reference Junction - As shown in Figure 18 and 18, two reference
junction configurations are possible: 1) reference located at the
datalogger or 2) reference at the AM16/32B.
Datalogger Reference - The CR1000, CR800, CR850, CR3000,
CR23X, 21X and the CR7 723-T Analog Input card with RTD have
built-in temperature references. The CR10XTCR Thermocouple
Reference (not standard with CR10X purchase), is installed on the
wiring panel between the two analog input terminal strips.
When the reference junction is located at the datalogger, the signal
wires between the data-logger and the AM16/32B must be of the
same wire type as the thermocouple (Figure 18). The "polarity" of the
thermocouple wires must be maintained on each side of the
multiplexer (e.g. if constantan wire is input to an L terminal, then a
constantan wire should run between the multiplexer's COM ODD L
terminal and the datalogger measurement terminal). Figures 17 and
18 depict type T thermocouple applications, but other thermocouple
types (e.g. E, J, and K) may also be measured and linearized by the
dataloggers.
If thermocouples are measured with respect to the datalogger
reference (i.e. the signal wires between datalogger and AM16/32B are
made of thermocouple wire), then it is not recommended that one
make measurements of any other sensor type through the AM16/32B.
31
Vista de pagina 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 49 50

Comentarios a estos manuales

Sin comentarios