AM16/32B Relay Analog Multiplexer
6.2 Differential Analog Measurement without Sensor
Excitation
Sensor to Multiplexer wiring - Up to two differential sensors that
don't require excitation may be connected to one input SET with panel
switch set to “4x16” mode. Sensor shields are connected to the input
” terminals.
Multiplexer to Datalogger wiring - The two pairs of COM terminals
(ODD H, ODD L and EVEN H, EVEN L) are connected to two pairs
of differential analog inputs at the datalogger. Observe H to H and L
to L from sensor to multiplexer to analog input. In “4x16” mode up to
32 differential sensors can be measured by two differential datalogger
channels in this way.
With panel switch set to “2x32” mode, one differential input can
measure up to 32 differential sensors in SETs of two with appropriate
programming.
6.3 Half Bridge Measurements
Measurements of this type may be subdivided into three categories
based on completion resistance and the presence or absence of
measured excitation. If the sensor's completion resistor(s) are
installed at the datalogger panel (example: a CSI 107 probe modified
for multiplexer use), then three probes per SET may be excited and
measured in “4x16” mode (Figure 13). However, if the circuit is
completed within the sensor (e.g. potentiometers), then excitation,
wiper signal, and ground must be multiplexed. Because excitation
and ground may be multiplexed in common, up to two sensors per
SET may be measured (Figure 14). If measured excitation is required
(i.e. four wire half-bridge), then only one sensor per SET of four may
be measured (Figure 15).
6.3.1 Half Bridge Measurement with Completion Resistor at
Datalogger
Sensor to Multiplexer wiring - up to three half bridges may be
connected to one input SET in “4x16” mode, provided that the
sensors’ three completion resistors are located at the datalogger
(Figure 13).
Multiplexer to Datalogger wiring - Signal lines from the multiplexer
COM terminals tie to three consecutive single-ended analog input
channels. Three precision completion resistors connect from analog
input channels to analog ground in CR10(X) or to “
” in the other
dataloggers.
27