CMP6-L, CMP10-L, CMP11-L, and CMP21-L Pyranometers
A-5
8. Attach the power cable to the CVF3 connector.
A.3.2 CVF4 Installation
To install, do the following:
1. Screw the CVF4 flat onto the mounting plate. Be sure to position the CVF4
so that the fan sticks out over the edge of the plate. See Figure A-7 for the
CVF4 mounting holes size and distances. The CVF4 is mounted with the
supplied washers and screws to the L31153 mount. Please note that 4 sets of
mounting hardware are included to mount the CVF4.
A-7 Mounted CVF4
2. Mount the radiometer in the CVF4 and level with its own levelling feet. The
radiometer mounting screws and shoulder washers are fitted (not secured).
If the CVF4 is mounted on a solar tracker, finish the alignment of
the traker first before securing the radiometer screws.
3. After all mounting screws have been secured, the CVF4 and radiometer
cables can be connected. The radiometer cable goes (down) through the slit in
the CVF4 base plate.
4. Put the CVF4 cover on; the 2 cover nuts can be fastened. Make sure the
cover is horizontal (check that there is equal space around the radiometer
dome) and that the 2 cover nuts are fastened hand tight.
Ø 7 mm (2x)
Vista de pagina 37