Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Sensor
10
The following instructions would trigger the agitator as discussed in the agitator
manual.
In the CR510, CR10X, and CR23X, use instruction P86 and P22.
EXAMPLE 8. Sample Program using P86 and P22 Instructions
45: Do (P86)
1: 41 Set Port 1 High
46: Excitation with Delay (P22)
1: 1 Ex Channel
2: 20 Delay W/Ex (units = 0.01 sec)
3: 0 Delay After Ex (units = 0.01 sec)
4: 0 mV Excitation
47: Do (P86)
1: 51 Set Port 1 Low
48: End (P95)
In the CR200 series dataloggers, use the Portset instruction.
EXAMPLE 9. Sample Program using Portset
Portset (1,1)
Delay(500,msec)
Portset(1,0)
In the CR1000, CR800, CR850, and CR3000 dataloggers, use the Portset
instruction.
EXAMPLE 10. Sample Program using Portset
Portset (1,1)
Delay (1,500,msec)
Portset (1,0)
The above example is not as power efficient as possible and would require AC
power to maintain a sufficient battery charge. If it is necessary to operate an
agitator without AC power available, write the program so that the agitator is only
operated for a short period of time just before the measurement is to be taken.