User Manual
3
3. Optional Sensor Accessories
009994 — PT4 Agitator for stagnant conditions
010315 — DO electrolyte, 1 litre
010314 — Membrane kit, 10 membranes
4. Optional Agitator – available to special order only
The PT4 Agitator is a reliable and robust agitator for use in conjunction with sensors
subjected to bio-fouling in ponds and stagnant water conditions.
O
2
sensors require a minimum water velocity across their membranes to function
properly. Therefore, to measure DO in stagnant water conditions, it is necessary to
move the water past the membrane to get accurate and reliable DO measurements. In
many instances the water also has a high bio-loading and the sensors become fouled
resulting in inaccurate DO measurements.
Figure 4-1. Preventing Bio-fouling of the CS512
The PT4 Agitator overcomes these problems. The device is designed so that a soft
bristle brush sweeps across the sensor membrane or sensor tip. This sweeping action
of the brush provides the required water velocity as well as prevents the membrane
from becoming bio-fouled.
The optimum sweeping frequency depends upon the design of sensor and type of
membrane used and water conditions. An ON-time of 0.25 seconds and OFF-time of
5 seconds is suitable in most circumstances.
Agitator’s overall size: 180mm x 83mm, diameter, 0.6 kg (7-1/8" x 3-1/4", 1-1/4 lb).
Supplied with 3 metres (10 ft.) cable.
Power required: 10.5 to 18 VDC at the agitator, 1.1 amps. Maximum ON-time is
3 seconds.