Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Sensor
4
Figure 4-2. CS512 with PT4 Agitator
Optional Repeat Cycle Timer for Agitators
: Reciprocating action may be controlled
by the optional solid state Repeat Cycle Timer. It sends 12 VDC pulses to the agitator
coil. The ON-time is 0.25 sec. The OFF-time is adjustable from 3 to 12 sec.; requires
supply voltage 10 to 17 VDC; housed in watertight cylinder 170 mm x 50 mm
diameter (6.6" x 2").
5. Application Information
SCWin users: This manual was written primarily for those whose
needs are not met by SCWin. Your procedure is much simpler: just
add the CS512 sensor (it’s in the water folder), save your program,
and follow the wiring shown in Step 2 of SCWIN.
6. Wiring
The CS512 sensor can use one differential channel or one single-ended channel.
Differential wiring is better at rejecting electrical noise and ground loop error.
Table 6-1. Sensor Wiring
Colour
Function
CR510, CR10X, CR800, CR850,
CR23X, CR1000, CR3000
CR200 Series
White Signal + Differential High,
or Single-Ended Channel
Single-Ended
Channel
Blue Signal - Differential Low or AG Ground
NOTE