User Manual
7
Table 7-2. Wiring for Example 2
CR200 Connection Sensor Wire
SE1 White
Ground Blue
Ground Clear
'CR200 Series
'Created by Short Cut (2.5 Beta)
'Declare Variables and Units
Public Batt_Volt
Public DOmV
Public DOppm
Units Batt_Volt=Volts
Units DOmV=mV
Units DOppm=ppm
'Define Data Tables
DataTable(Table1,True,-1)
DataInterval(0,60,Min)
Sample(1,DOmV)
EndTable
DataTable(Table2,True,-1)
DataInterval(0,1440,Min)
Minimum(1,Batt_Volt,False,False)
EndTable
'Main Program
BeginProg
Scan(10,Sec)
'Default Datalogger Battery Voltage measurement Batt_Volt:
Battery(Batt_Volt)
'CS512 Dissolved Oxygen Probe measurements DOmV and DOppm:
VoltSE(DOmV,1,1,1,0)
DOppm=DOmV*0.34 'Note use exact calibration factor (see section 8)
'Call Data Tables and Store Data
CallTable(Table1)
CallTable(Table2)
NextScan
EndProg