SECTION 1. FUNCTIONAL MODES
1-15
TABLE 1.8-10. Set Initial Baud Rate / Set
RS232 Power
Key
Entry Display
Comments
D
13:Enter Command
00
1
2
A
12: Connect Baud Rate
00
X
C
A
12: Connect Baud Rate Enter Baud
0X-- Rate Code X
(Table 1.8-11).
Index (--)
is optional.
TABLE 1.8-11. Baud Rate Codes
X = 0 300 Baud
X = 1 1200 Baud
X = 4 2400 Baud
X = 5 4800 Baud
X = 2 9600 Baud
X = 6 19.2 K Baud
X = 7 38.4 K Baud
X = 3 76.8 K Baud
X-- = RS232 Power On
1.8.8 SET PROGRAM COMPILE OPTION
Table 1.8-13 shows the option codes available
for setting the program compile option. This
setting will affect the program compile when the
program is downloaded from the PC or a
SM192/716 Storage Module. It also affects
compiling with
B
and the arcane
D
7
command. Keyboard or Remote Keyboard
compiling with
6
and
0
is not
affected by this setting. If a .DLD file has this
setting, it will affect the compile operation
AFTER the .DLD file is downloaded.
TABLE 1.8-12. Set Program Compile Option
Key
Entry Display
Comments
D
13:Enter Command
00
1
3
A
13: Compile Option
00
1
A
13: Compile Option Sets Compile
01 like
6
TABLE 1.8-13. Compile Option Codes
0 Compile like
0
(See Section 1.4)
1 Compile like
6
(See Section 1.3)
2 Do not clear intermediate storage
Vista de pagina 48