Model HFP01SC Self-Calibrating Soil Heat Flux Plate
;Begin heating for calibration.
;
6: Do (P86)
1: 48 Set Port 8 High
;Used to filter data during and after calibration.
;
7: Do (P86)
1: 18 Set Flag 8 High
8: End (P95)
;End site calibration three minutes after calibration started.
;
9: If time is (P92)
1: 4 Minutes (Seconds --) into a
2: 180 Interval (same units as above)
3: 30 Then Do
;Measure voltage across current shunt resistor during calibration.
;This measurement is used to compute power.
;
10: Volt (SE) (P1)
1: 6 Reps
2: 25 2500 mV 60 Hz Rejection Range
3: SE Channel
4: 37 Loc [ V_Rf_1 ]
5: .001 Mult
6: 0 Offset
11: Beginning of Loop (P87)
1: 0 Delay
2: 6 Loop Count
12: Z=X (P31)
1: 7 -- X Loc [ shf_mV_1 ]
2: 25 -- Z Loc [ mV_180_1 ]
13: End (P95)
;Turn off the soil heat flux plate heaters.
;
14: Do (P86)
1: 58 Set Port 8 Low
15: End (P95)
;Compute in-situ calibration.
;
16: If time is (P92)
1: 39 Minutes (Seconds --) into a
2: 180 Interval (same units as above)
3: 30 Then Do
21