User Manual
7.2.3.4 SplitStr() Instruction
The SplitStr() instruction has the following syntax:
SplitStr(SplitResult,SearchString,FilterString,NumSplit,SplitOption)
SplitResult — an array in which the split string will be stored.
SearchString — the string on which this instruction will operate. This will be
the variable entered for the Dest parameter for the SerialIn() instruction
(see above).
FilterString — enter “String” (this value will be ignored because of the
SplitOption that will be used).
NumSplit — enter 2
SplitOption — enter 0. This splits out numeric values.
7.2.4 Programming for Calibration
To output in pH units instead of digital counts, enter the offset and multiplier into
the datalogger program. Simple program instructions can be used to make the
required periodic calibration easier. See Appendix C, Calibration, for an example
program.
7.3 Calibration
Calibration should be carried out according to the detailed procedure outlined in
Appendix C, Calibration. The calibration should use two or more pH standards,
listed in Table 7-4, which are available from Campbell Scientific.
Table 7-4. Calibration Standards
pH Part Number
4 #25587
7 #25586
10 #25588
Frequency of calibration depends on the level of accuracy required and the
coating / fouling nature of the measured samples.
8. Troubleshooting
Contact Campbell Scientific if the CS526 is not operating properly
and requires return to Campbell Scientific Ltd.
The most common causes for erroneous pH data include:
• poor sensor connections to the datalogger
• damaged cables
• scratched chip
• contaminated or clogged diaphragm