Campbell CS526 Especificaciones Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 30
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
CS526 ISFET pH Probe
Sensor Material: PEEK
Weight
w/10-ft cable: 318 g (11.2 oz)
Dimensions
Length: 102 mm (4 in)
Diameter: 16 mm (0.63 in)
Certifications: ISO 9001
compliant
EMC standard IEC61326:2005
7. Operation
If you are programming your datalogger with Short Cut, skip Section 7.1, Wiring,
and Section 7.2, Datalogger Programming. Short Cut does this work for you.
See Section 4, Quickstart, for a Short Cut tutorial.
7.1 Wiring
Table 7-1. Wiring
Wire Colour Wire Label/Function Datalogger Connections
Red (see following
caution)
Power 5V 5V
Black Ground G
White Signal #1 (Rx) Control Port (Tx)
Green Signal #2 (Tx) Control Port (Rx)
Clear Shield G
This probe must be connected to the datalogger’s 5 V
terminal (not 12 V). Connecting to a higher voltage will
damage the probe beyond repair.
7.2 Datalogger Programming
7.2.1 Get Data Command
The datalogger needs to send a “get data” serial command to the CS526 to get the
pH data. This command is sent to the CS526 via the SerialOut() CRBasic
instruction (see Section 7.2.3.2, SerialOut() Instruction). Table 7-2 shows the “get
data” command and its response.
CAUTION