CS451/CS456 Submersible Pressure Transducer
The desiccant in the tube changes color from blue to pink when the drying
power is lost.
The Enclosure Accessory Humidity Indicator Card (pn 28878).
8.1 Every Visit
Collect data
Visually inspect wiring and physical conditions
Check indicating desiccant or enclosure humidity indicator; service if
necessary
Check battery condition (inspect physical appearance and use a keyboard
display, PDA, or laptop to view the battery voltage)
Check all sensor readings: adjust transducer offsets if necessary
Check recent data
8.2 Every Two to Three Years or on a Rotating Schedule
Send the CS451/CS456 in for inspection.
9. Troubleshooting
The most common causes for erroneous pressure transducer data include:
poor sensor connections to the datalogger
damaged cables
damaged transducers
moisture in the vent tube
Problem:
Unit will not respond when attempting serial communications.
Suggestion:
Check the power (red is +V and black is ground) and signal (white is SDI-12
data) lines to ensure proper connection to the datalogger. Check the datalogger
program to ensure that the same port the SDI-12 data line is connected to is
specified in the measurement instruction.
Problem:
Transducer appears to be operating properly but data shows a periodic or cyclic
fluctuation not attributable to water level changes.
24