CS106 Barometric Pressure Sensor
1. General
The CS106 analog barometer uses Vaisala’s Barocap® silicon capacitive
pressure sensor. The Barocap sensor has been designed for accurate and stable
measurement of barometric pressure. The CS106 outputs a linear 0 to
2.5 VDC signal that corresponds to 500 to 1100 mb. It can be operated in a
shutdown or normal mode. In the shutdown mode the datalogger switches
12 VDC power to the barometer during the measurement. The datalogger then
powers down the barometer between measurements to conserve power.
2. Specifications
Operating Range
Pressure: 500 mb to 1100 mb
Temperature: -40°C to +60°C
Humidity: non-condensing
Accuracy
Total Accuracy*** ±0.3 mb @ +20°C
±0.6 mb @ 0°C to +40°C
±1 mb @ -20°C to +45°C
±1.5 mb @ -40°C to +60°C
Linearity*: ±0.25 mb @ 20°C
Hysteresis*: ±0.03 mb @ 20°C
Repeatability*: ±0.03 mb @ 20°C
Calibration Uncertainty**: ±0.15 mb @ 20°C
Long-Term Stability: ±0.1 mb per year
* Defined as ±2 standard deviation limits of end-point non-linearity,
hysteresis error, or repeatability error
** Defined as ±2 standard deviation limits of inaccuracy of the working
standard at 1000 mb in comparison to international standards (NIST)
*** Defined as the root sum of the squares (RSS) of end-point non-linearity,
hysteresis error, repeatability error and calibration uncertainty at room
temperature
General
Dimensions: 9.7 cm x 6.8 cm x 2.8 cm (3.8” x 2.7” x 1.1”)
Weight: 90 g (3.2 oz)
Housing Material: ABS/PC blend
Supply Voltage: 10 to 30 VDC
1