Campbell CNR4 Manual de instrucciones Pagina 29

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 70
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 28
CNR4 Net Radiometer
FIGURE 7-2. The marks on the end of the CNR4: S for SOLAR cable,
and T for TEMP cable
The measurement details for Pt-100 sensor, including the wiring diagram and
sample program, are explained in Appendix B.4, CR3000 Program for
Measuring Pt-100 Temperature Sensor.
The four radiation outputs can be measured using differential or single-ended
inputs on the datalogger. A differential voltage measurement is recommended
because it has better noise rejection than a single-ended measurement.
When differential inputs are used, jumper the low side of the input
to AG or to keep the signal in common mode range.
TABLE 7-3 and TABLE 7-4 show the wiring instructions for the differential
measurement and single-ended measurement connections to the datalogger,
respectively. The cables have the white band at the pigtail end of the cable
with the color keys. See FIGURE 7-3 and FIGURE 7-4 below for the labels on
the cable for both the SOLAR and TEMP cables.
FIGURE 7-3. Labels on the pigtail end of the SOLAR cable
NOTE
19
Vista de pagina 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34 ... 69 70

Comentarios a estos manuales

Sin comentarios