CNR1 Net Radiometer
Pt-100 Temperature Sensor Connections to 4WPB100 and Datalogger
Function Color 4WPB100 Datalogger
Black Wire Excitation
H Differential Input (H)
Pt-100 Excitation + Red L Differential Input (L)
Pt-100 Excitation - Blue G (AG CR10X/CR510)
Pt-100 Signal + Yellow Differential Input (H)
Pt-100 Signal – Green Differential Input (L)
TABLE 5-2. Datalogger Connections for Single-Ended Measurement, When Using a 4WPB100
Function Color CR10X,CR510 CR23X 21X/CR7
CM3 Up Signal Red Single-Ended Input Single-Ended Input Single-Ended Input
CM3 Up Reference Blue AG
CM3 Down Signal White Single-Ended Input Single-Ended Input Single-Ended Input
CM3 Down Reference Black AG
CG3 Up Signal Grey or Orange Single-Ended Input Single-Ended Input Single-Ended Input
CG3 Up Reference Yellow AG
CG3 Down Signal Brown Single-Ended Input Single-Ended Input Single-Ended Input
CG3 Down Reference Green AG
Shield Shield G
TABLE 5-3. CR3000 and CR5000 Connections
for Differential Measurement
Function Color CR3000/CR5000
CM3 Up Signal Red Differential Input (H)
CM3 Up Reference Blue Differential Input (L)
CM3 Down Signal White Differential Input (H)
CM3 Down Reference Black Differential Input (L)
CG3 Up Signal Grey or Orange Differential Input (H)
CG3 Up Reference Yellow Differential Input (L)
CG3 Down Signal Brown Differential Input (H)
CG3 Down Reference Green Differential Input (L)
PT-100 Signal + Yellow Differential Input (H)
PT-100 Signal - Green Differential Input (L)
PT-100 Current Excitation + Red Current Excitation IX
PT-100 Current Excitation - Blue Current Excitation IXR
Shield Clear
6. Datalogger Programming
The CNR1 outputs four voltages that typically range from 0 to 15 mV for the
CM3 sensors, and ± 5 mV for the CG3 sensors. A differential voltage
measurement (Instruction 2) is recommended because it has better noise
rejection than a single-ended measurement. If differential channels are not
available, single-ended measurements (Instruction 1) can be used. The
acceptability of a single-ended measurement can be determined by simply
10