CNR1 Net Radiometer
EX1
4WPB100
Red
CNR1
DATALOGGER
Yellow
Green
PT-100
Blue
H
L
G
10K
5H
5L
6H
6L
100
FIGURE 5-2. Interfacing the Pt-100 Using the 4WPB100 Module
The four radiation outputs can be measured using Differential or Single-Ended
inputs on the datalogger. A differential voltage measurement (Instruction 2) is
recommended because it has better noise rejection than a single-ended
measurement. When differential inputs are used, jumper the low side of the
input to AG or
to keep the signal in common mode range.
TABLE 5-1. Datalogger Connections for Differential Measurement, When Using a 4WPB100
Function Color CR10X,CR510 CR23X 21X/CR7
CM3 Up Signal Red Differential Input (H) Differential Input (H) Differential Input (H)
CM3 Up Reference *Blue Differential Input (L) Differential Input (L) Differential Input (L)
CM3 Down Signal White Differential Input (H) Differential Input (H) Differential Input (H)
CM3 Down Reference *Black Differential Input (L) Differential Input (L) Differential Input (L)
CG3 Up Signal Grey or
†Orange
Differential Input (H) Differential Input (H) Differential Input (H)
CG3 Up Reference *Yellow Differential Input (L) Differential Input (L) Differential Input (L)
CG3 Down Signal Brown Differential Input (H) Differential Input (H) Differential Input (H)
CG3 Down Reference *Green Differential Input (L) Differential Input (L) Differential Input (L)
Shield Shield G
*Jumper to AG or with user supplied wire.