Campbell Scientific TE525-L Tipping Bucket Rain Gauge
RM72004,0000165 -19-07MAY13-1/1
Troubleshooting
NOTE: If troubleshooting indicates sensor damage,
contact your John Deere dealer.
Verify:
power to sensor
correct connections
tight connections
no kinks in harness
Symptom: No Precipitation
NOTE: Resistance measured at terminal block (on inside
of bucket between black and white leads) must be
less than 1 Ohm when bucket is balanced.
Resistance varies when bucket is tipped.
1. Disconnect sensor; use Ohm meter to check switch
for continuity.
2. Check connection at gateway.
70-2
052813
PN=58