TABLE 10-1. Output Modes for EC100 Instruction
Output
Mode
Data
Field
Description
Units
1 Ux m/s
2 Uy m/s
3 Uz m/s
4 Sonic Temperature ºC
5 Sonic Diagnostic Flag
6
CO
2
Density mg/m
3
7 H
2
O Density g/m
3
0, 1, 2,
8 Gas Diagnostic Flag
9 Air Temperature ºC
10 Air Pressure kPa
11
CO
2
Signal Strength
nominally 0.0 ≤ strength
≤1.0
1, 2
12
H
2
O Signal Strength
nominally 0.0 ≤ strength
≤1.0
2 13
Sample Cell Pressure
Differential
kPa
As shown in Table 10-1, all output modes give two diagnostic values: Sonic
Diagnostic Flag and Gas Diagnostic Flag. The values contain a bit field with
each bit representing a monitored condition. When a certain condition is
detected, the corresponding bit is set. The value remains set until the event that
caused the condition is no longer present. Tables 10-2 and 10-3 describe the
bits in the sonic diagnostic flag and the gas diagnostic flag, respectively.
TABLE 10-2. Bits in the Sonic Diagnostic Flag
bit
hex
value
decimal
Name
Function
0 0x1 1 Low Amp
Acoustic signal has
dropped below threshold
1 0x2 2 High Amp
High-amplitude threshold
exceeded
2 0x4 4 Tracking Tracking error
3 0x8 8 Hi 3 Axis DC
Measured speed of sound
exceeds threshold
4 0x10 16 Acquiring
Sonic is re-acquiring
signals
5 0x20 32 Cal Mem Err
Head calibration memory
signature error