Appendix C. HC2S3 Digital Communications
C-3
C.3 RS-485 Communications using an MD485 RS-485
Interface
The HC2S3 can be interfaced to a CSI datalogger through an MD485 RS-485
Interface using the Rotronic E2-05XX-MOD RS485 cable as described below.
Settings for the RS485 port on the MD485 must be configured to match the
configuration of the HC2S3, which are 19200 baud, No Parity, 8 Data Bits,
1 Stop bit, and No Flow Control.
Device Configuration Utility (CSI software available as a free download) is used
to configure the MD485. Configuration settings for the MD485 are shown
below:
MD485 Tab: CS I/O AND RS-485
CS I/O Tab: SDC Address 7
RS485 Tab: RS485 baud 19200