Campbell HC2S3 Manual Pagina 19

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 36
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 18
Model HC2S3 Temperature and Relative Humidity Probe
CR1000 program using differential measurements
'CR1000 program to measure HC2S3 with differential measurements
Public AirTC
Public RH
DataTable(Temp_RH,True,-1)
DataInterval(0,60,Min,0)
Average(1,AirTC,IEEE4,0)
Sample(1,RH,IEEE4)
EndTable
BeginProg
Scan(1,Sec,1,0)
'HC2S3 Temperature & Relative Humidity Sensor measurements AirTC and RH:
VoltDiff (AirTC,1,mV2500,1,True,0,_60Hz,0.1,-40)
VoltDiff (RH,1,mV2500,2,True,0,_60Hz,0.1,0)
If RH>100 And RH<103 Then RH=100
CallTable(Temp_RH)
NextScan
EndProg
CR10(X) program using differential measurement instructions
;{CR10X}
*Table 1 Program
01: 1.0000 Execution Interval (seconds)
;HC2S3 Temperature & Relative Humidity Sensor measurements AirTC and RH:
1: Volt (Diff) (P2)
1: 1 Reps
2: 25 2500 mV 60 Hz Rejection Range
3: 1 DIFF Channel
4: 3 Loc [ AirTC ]
5: 0.1 Multiplier
6: -40 Offset
2: Volt (Diff) (P2)
1: 1 Reps
2: 25 2500 mV 60 Hz Rejection Range
3: 2 DIFF Channel
4: 4 Loc [ RH ]
5: 0.1 Multiplier
6: 0 Offset
3: If (X<=>F) (P89)
1: 4 X Loc [ RH ]
2: 3 >=
3: 100 F
4: 30 Then Do
13
Vista de pagina 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 35 36

Comentarios a estos manuales