
CS526 ISFET pH Probe
4. Quickstart
For complete installation, programming, and calibration information, see
Sections 7 through 9.
The CS526 is shipped dry. Before the probe can be calibrated, it must be
soaked in 7 pH buffer solution for 15 minutes. Follow the calibration
procedure outlined in Appendix A.
Unlike many glass-bulb type pH sensors, the CS526 ISFET pH probe can be
installed without regard to orientation. Connect the CS526 to the datalogger
according to the following table. Be sure to connect Red to the 5V channel.
Wire Color Datalogger Connection
Red (Caution! 5 Vdc Max!) 5V
Black G
White Control Port (Tx)
Green Control Port (Rx)
Clear G
Program the datalogger as follows.
'CR1000 Series Datalogger
'Declare variable for PH measurement in mV
Public pHmV
'Declare variables for serial input from sensor
Dim rawstring As String * 20, pHDigit(2)
'Main Program
BeginProg
'Set up one of the datalogger's ports for serial communication
SerialOpen ( Com1,2400,16,0,41)
'Establish program scan rate of 60 seconds
Scan (60,Sec,0,0)
'Send get data command to CS526
SerialOut (Com1,"1M1!"&CHR(13),"",0,0)
'Set up COM1 to receive incoming serial data.
'Set timeout to maximum 200mS
SerialIn (rawstring,Com1,20,0,20)
'Split out numeric mV value for PH from string input
SplitStr (pHDigit(),rawstring,"String",2,0)
pHmV = pHDigit(2)
NextScan
EndProg
2
Comentarios a estos manuales