CS526 ISFET pH Probe
11. Maintenance
Campbell Scientific recommends that the CS526 be cleaned and calibrated
periodically to insure accurate readings.
The CS526 has no user-serviceable parts.
If the CS526 is not operating properly and requires return to Campbell
Scientific, first obtain an RMA (returned material authorization) and fill out
the Declaration of Hazardous Material and Decontamination form. This
information is available at http://www.campbellsci.com/repair
.
Cable can be damaged by abrasion, rodents, sharp objects, twisting, crimping
or crushing and pulling. Take care during installation and use to avoid cable
damage.
Contamination on the diaphragm may block the built-in reference electrode, so
there will be no electrical contact between the electrode and the ISFET chip.
This condition will cause the probe to not function. To prevent this, regular
cleaning is recommended. Frequency of cleaning depends on the quality of the
water. The best cleaning method is the “soapy water” method:
1. Place the tip of the probe in soapy water that is 50° to 60°C and soak for
about 5 minutes (see Figure 11-1).
2. Rinse thoroughly with deionized water.
3. If necessary, lightly scrub the tip of the probe with a toothbrush to remove
debris (see following caution).
4. Place the probe in saturated KC1 solution (CSI pn 16349) at about 20°C.
Keep probe in solution for about 30 minutes.
To prevent scratching the chip, a toothbrush should only
be used after soaking in soapy water. Most scratches
occur when hard particles are rubbed with a toothbrush on
the chip surface. This can cause irreversible damage to
the probe.
CAUTION
5. Rinse the probe.
6. Calibrate the probe (see Appendix A).