CDM-VW300 Series Dynamic Vibrating-Wire Analyzers
FIGURE 6-13. Earth ground connections
9. Write or obtain a CRBasic program for the datalogger. Reference Section
7.10, CRBasic Programming, and Appendix G, CRBasic Program
Library.
CDM-VW300 and CDM-VW305 programs use the following CRBasic
instructions:
CDM_VW300Config()
CDM_VW300Dynamic()
CDM_VW300Static()
The CRBasic program must be enabled specifically for the CDM-VW300
or the CDM-VW305. Programs enabled for a CDM-VW300 use a
DeviceType argument of 0 in the CDM_VW300Config() instruction.
Programs enabled for a CDM-VW305 use a DeviceType of 1. Check that
other settings recorded in step 2 are integrated into the CRBasic program.
Do not attempt to use a program written for the CDM-VW300
when using a CDM-VW305. Configurations for a particular
analyzer are deeply rooted in the CRBasic program. Simply
changing the DeviceType argument will not make all the
necessary changes.
NOTE
24