CR10 and VW Sensors, 2002/12/18 6
Check for Reading Quality
Instruction 131 Only
Instruction 131 instruction returns two frequency values in
consecutive locations. The first value, Freq_1, is produced from
the full number of cycles. The second value, Freq_2, is produced
from the first 20% of the full number of cycles.
The deviation between the two frequencies is a measure of read-
ing quality. Deviation is defined as:
The maximum allowable deviation for each sweep range is
shown in the table below. If the deviation is less than the maxi-
mum, the Freq_1 value can be accepted as a good reading. If the
deviation is greater, the reading should be taken again. (Always
provide a delay of at least two seconds before repeating a read-
ing).
Sweep Range Maximum Allowable Deviation
10.008
20.008
30.0032
40.0016
Deviation ABS 1
Freq_2
Freq_1
---------------- -–
=