CS450/CS455 Pressure Transducer
3.1.3 SDI-12 Transparent Mode
Transparent Mode allows direct communication with the CS450/CS455. This
may require waiting for programmed datalogger commands to finish before
sending responses. While in the transparent mode, datalogger programs may
not execute. Datalogger security may need to be unlocked before transparent
mode can be activated.
Transparent mode is entered while the PC is in telecommunications with the
datalogger through a terminal emulator program. It is most easily accessed
through Campbell Scientific datalogger support software, but is also accessible
with terminal emulator programs such as Windows HyperTerminal.
To enter the SDI-12 transparent mode, enter Terminal Emulator from
LoggerNet, PC400 or PC200W datalogger support software. A terminal
emulator screen is displayed. Click the “Open Terminal” button as shown in
Figure 3-5.
For CR800 series, CR1000, CR3000 loggers, press <Enter> until the
datalogger responds with the prompt (“CR1000>” for the CR1000). Type
“SDI12” at the prompt (without quotes) and press <Enter>. In response, the
query “Enter Cx Port 1,3,5 or 7” will appear. Enter the control port integer to
which the CR450 is connected. An “Entering SDI12 Terminal” response
indicates that SDI-12 Transparent Mode is active.
FIGURE 3-5. Terminal Emulator
For CR10X and CR510 standard mixed-array loggers: The datalogger
telecommunications command to enter SDI-12 transparent mode is
“nX<Enter>” where n is the control port being used for SDI-12. In this
example the selected control port is C1, so the command would be
“1X<Enter>”. In response, the datalogger opens the link to control port 1 and
8