Vista de pagina 4
CS15-L Current Transformer
1. General Description
The CS15-L uses a modified CR Magnetic’s CR8459 Current Transformer to
measure the approximate current over a range of 0 to 125 A. The CS15-L
outputs a millivolt signal allowing it to be directly connected to the CR2XX
dataloggers. The CS15-L is recommended for measurements that do not
require high accuracy.
FIGURE 1. CS15-L Current Transformer
2. Specifications
Example Applications:
Motor or generator load conditions
Efficiency studies
Intermittent fault detection
Rough submetering
Manufacturer’s Specifications
Frequency: 50 and 60 Hz
Insulation Resistance: 100 M ohm @ 500 VDC
High Potential: 2000 volts
Rated Current: 125 A
Storage Temperature: -25ºC to 70ºC
Operating Temperature: -25ºC to 55ºC
Case Material: Polypropylene Resin
Construction: Epoxy Encapsulated
Accuracy with 10 ohm
burden max. (resistive): typically ±5 percent of actual value with
provided multiplier
Dimensions: Outer diameter: 1.89” (4.8 cm)
Inner diameter: 0.75” (1.9 cm)
Depth: 0.67” (1.7 cm)